• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC PUERTO RICO OPERATIONS CO. 530G INSULIN PUMP MMT-551NAB; ARTIFICIAL PANCREAS DEVICE SYSTEM, THRESHOLD SUSPEND

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC PUERTO RICO OPERATIONS CO. 530G INSULIN PUMP MMT-551NAB; ARTIFICIAL PANCREAS DEVICE SYSTEM, THRESHOLD SUSPEND Back to Search Results
Model Number MMT-551NAB
Medical Device Problem Code Failure to Cycle (1142)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 05/29/2020
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).Currently it is unknown whether or not the device may have caused or contributed to the event as no product has been returned.The device will be returned for analysis and further information will follow once the analysis has been completed.No conclusion can be drawn at this time.
 
Event or Problem Description
Information received by medtronic indicated that the insulin pump had compromised force sensor system alarm.Customer stated insulin pump was able to clear alarm however not able to rewind the alarm.No harm requiring medical intervention was reported.The device will be returned for analysis.
 
Additional Manufacturer Narrative
Device passed the displacement test.Device received with motor error alarm during the basic occlusion test due to loose or protruded drive support disk.Unable to perform prime/a33 test, excessive no delivery test and occlusion test or verify a33 alarm due to motor error alarm.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
530G INSULIN PUMP MMT-551NAB
Common Device Name
ARTIFICIAL PANCREAS DEVICE SYSTEM, THRESHOLD SUSPEND
Manufacturer (Section D)
MEDTRONIC PUERTO RICO OPERATIONS CO.
ceiba norte ind. park #50 road
juncos 00777 -386
MDR Report Key10122343
Report Number2032227-2020-126660
Device Sequence Number2341964
Product Code OZO
UDI-Device Identifier00643169503625
UDI-Public(01)00643169503625
Combination Product (Y/N)N
Initial Reporter StateMI
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source consumer
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 06/24/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Model NumberMMT-551NAB
Device Catalogue NumberMMT-551NAB
Device Lot NumberA4551NABJ
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/09/2020
Initial Date Received by Manufacturer 05/29/2020
Supplement Date Received by Manufacturer06/24/2020
Initial Report FDA Received Date06/05/2020
Supplement Report FDA Received Date06/24/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age40 YR
Patient Weight190
-
-