• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: C.R. BARD, INC. (COVINGTON) -1018233 BARD TOUCHLESS PLUS UNISEX PRE-LUBRICATED VINYL URETHRAL CATHETER; TOUCHLESS 16FR CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

C.R. BARD, INC. (COVINGTON) -1018233 BARD TOUCHLESS PLUS UNISEX PRE-LUBRICATED VINYL URETHRAL CATHETER; TOUCHLESS 16FR CATHETER Back to Search Results
Model Number 4A5146
Medical Device Problem Code Inaccurate Flow Rate (1249)
Health Effect - Clinical Codes No Consequences Or Impact To Patient (2199); No Known Impact Or Consequence To Patient (2692)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The investigation is still in progress.Once the investigation is complete a supplemental report will be filed.The device was not returned.
 
Event or Problem Description
It was reported that the kit did not work well while draining the bladder when on bed.
 
Additional Manufacturer Narrative
The reported event was inconclusive, as no sample returned for evaluation.A potential root cause for this failure mode could be due to ¿poor interface with connections.¿ it was unknown whether the device had met specifications.The product used for treatment purposes.It was unknown whether the product had caused the reported failure.The device was not returned for evaluation.The lot number was unknown, therefore, the device history record could not be reviewed.The instructions for use were found adequate and state the following: "caution: federal (usa) law restricts this device to sale by or on the order of a physician.After use, this product may be a potential biohazard.Handle and dispose of in accordance with accepted medical practice and with applicable local, state and federal laws and regulations.Instructions for use: preparing for catheterization: 1.Open kit and remove contents.2.The gloves are not necessary to maintain aseptic conditions during the procedure.The gloves are for the operator's protection.Male/female catheterization: 1.Remove cap from insertion tip while squeezing tabs.Save cap for closing the bag.2.Slide the catheter halfway into the insertion tip.3.Male - prepare the male urethral meatus and the surrounding area with povidone-iodine swabs provided.Female - prepare the female urethral meatus by holding the outer labia apart and prep the urethral meatus and surrounding area with povidone iodine swabs provided.4.Male - holding the penis, advance the insertion tip into the urethra no further than the flange base.Female - spread the inner labia, advance the insertion tip into the urethra no further than the flange base.Release the inner labia.5.Place your nondominant hand on the finger control guide to stabilize catheter in urethra.With your dominant hand, grasp catheter through bag approximately 1" below the finger control guide and push catheter into urethra.The catheter should be introduced by short, repetitive pushing motions.Repeat motions until catheter reaches bladder and urine starts to flow.6.Allow urine to flow freely, making certain the catheter guide is elevated at least 4" above lower portion of the bag.Allow flow until bladder is empty or until bag is filled.Precaution: it is recommended that the collection bag be held.The catheter could possibly separate from the collection bag when urine increases weight of the bag.7.Withdraw the catheter from the urethra.Remove the remaining portion of the catheter from the collection bag by pulling it through the insertion tip.Note: the catheter is designed to pass through the insertion tip.8.The filled collection bag may be closed by replacing the cap over the insertion tip.Make sure the cap snaps on securely.Specimen collection: to collect a specimen, obtain a sample cup/tube and alcohol wipe.Remove guide tip by pulling tab(s) upward from bag.Wipe port with alcohol.Pour specimen through port at top of bag into cup/tube.Draining bag: remove insertion tip by pulling tab(s) upward from bag.Drain urine through port at top of bag.".
 
Event or Problem Description
It was reported that the kit did not work well while draining the bladder when on bed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BARD TOUCHLESS PLUS UNISEX PRE-LUBRICATED VINYL URETHRAL CATHETER
Common Device Name
TOUCHLESS 16FR CATHETER
Manufacturer (Section D)
C.R. BARD, INC. (COVINGTON) -1018233
8195 industrial blvd
covington GA 30014
MDR Report Key10122904
Report Number1018233-2020-03642
Device Sequence Number13976777
Product Code FCM
UDI-Device Identifier00801741046438
UDI-Public(01)00801741046438
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
PMA/510(K) Number
K910653
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source consumer,distributor,other
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 07/21/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Model Number4A5146
Device Catalogue Number4A5146
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 05/15/2020
Supplement Date Received by Manufacturer06/30/2020
Initial Report FDA Received Date06/05/2020
Supplement Report FDA Received Date07/21/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-