Event or Problem Description
It was reported that the procedure was cancelled.Vascular access was obtained utilizing a brachial arm approach.The target lesion was located in the right superficial femoral artery.An angiojet solent omni was used for a thrombectomy procedure.During procedure, it was noted that the tip of the catheter was stuck on the patient's implanted stent graft while the physician was advancing the device.The device was completely removed from the patient's body and found out that the tip looked frayed.The physician wanted to open another catheter but an angiojet solent dista was expired on the shelf.It was unable to proceed the treatment.No patient complications were reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ANGIOJET SOLENT DISTA
Common Device Name
CATHETER, EMBOLECTOMY
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
two scimed place
maple grove MN 55311
MDR Report Key10123128
Report Number2134265-2020-07568
Device Sequence Number2379509
Product Code DXE
UDI-Device Identifier08714729889731
UDI-Public08714729889731
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
PMA/510(K) Number
K113363
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 06/05/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date03/30/2020
Device Model Number45030
Device Catalogue Number45030
Device Lot Number0022151133
Was Device Available for Evaluation? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 05/21/2020
Initial Report FDA Received Date06/05/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-