| Model Number |
801-07545 |
| Medical Device Problem Codes |
Insufficient Device Problem Information (3190); Migration (4003)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
02/04/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
During a planned revision surgery to perform additional compression on an existing construct, the surgeon noted that a mesa polyaxial screw was not secured to the rod.Surgery was successfully completed with no delay and no adverse consequences to the patient were reported.
|
| |
|
Additional Manufacturer Narrative
|
|
Visual, dimensional, material and functional analysis could not be performed as the device was not returned.Device and complaint history records were reviewed for this lot, and no relevant manufacturing issues or similar complaints were identified.Post-operative patient activities may have introduced unanticipated loading to the construct, overloading the capture strength of the screw at l5.There is also a possibility that the subject screw was never locked during the index surgery.However, as the device remains implanted and unavailable for evaluation, the root cause could not be determined conclusively.H3 other text : device remains implanted.
|
| |
|
Event or Problem Description
|
|
During a planned revision surgery to perform additional compression on an existing construct, the surgeon noted that a mesa polyaxial screw was not secured to the rod.The device remains implanted.Surgery was successfully completed with no delay and no adverse consequences to the patient were reported.
|
| |
|