| Catalog Number |
H93863 |
| Medical Device Problem Codes |
Fluid/Blood Leak (1250); Material Split, Cut or Torn (4008)
|
| Health Effect - Clinical Code |
No Patient Involvement (2645)
|
| Date of Event |
05/13/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(6).Should additional relevant information become available, a supplemental report will be submitted.
|
| |
|
Event or Problem Description
|
|
It was reported that at least 1 trifurcated fluid transfer tube set leaked from a tear on the tubing which was threaded around the rotor.It was further reported that the pump was setup up with the port of the tubing connected to a 1l 0.9% normal saline bag and the outlet end of the tubing was setup with an 18 gauge needle.The pump setting was on high 7.This issue was identified during preparation of an empty bag.There was no patient involvement.No additional information is available.
|
| |
|
Additional Manufacturer Narrative
|
|
The lot was manufactured from july 11, 2018 - july 12, 2018.The device was received for evaluation.Visual inspection was performed which observed leakage and tear in the section of the tubing that wraps around the rotor.A functional testing was performed which revealed a leak on the affected area.The reported condition was verified.The cause of the condition was not determined.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.Should additional relevant information become available, a supplemental report will be submitted.
|
| |
|