| Catalog Number |
7207675 |
| Medical Device Problem Code |
Detachment of Device or Device Component (2907)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
05/12/2020
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported that during a cl reconstruction surgery the screw fractured when screw in, all pieces were removed and no additional bone holes were performed.The procedure was successfully completed with a backup device, no significant delay reported.All available information has been disclosed.If additional information should become available, a supplemental report will be submitted accordingly.
|
| |
|
Additional Manufacturer Narrative
|
|
H3, h6: two 7207675 scr biorci-ha 7x25 8mm head sterile intended for use in treatment have been returned for evaluation.The information provided states: ¿during a cl reconstruction surgery two screws were fractured when screw in, all pieces were removed and no additional bone holes was performed¿.Visual assessment confirms the complaint.The screws were returned fractured.An exact root cause cannot be determined with confidence; however, factors that could have contributed to the reported event include: excessive force applied during insertion.Improper site preparation.The instruction for use outlines precautionary statements and instructions in regards to the use of the device to avoid damage or non-functionality.There were no indications that would suggest that the device did not meet product specifications upon release into distribution.A review of the complaint and manufacturing records was performed and indicated similar allegations for the lot number reported.
|
| |
|