Model Number URF-P6R
Device Problem Break (1069)
Patient Problem No Patient Involvement (2645)
Event Type  malfunction  
Manufacturer Narrative
The device was returned to olympus for evaluation.The device was visually inspected and found the bending section cover and the skeleton is noted to be detached/broken.In addition, there were excessive broken fibers on the image due to the broken bending section skeleton.The instruction manual states the following; ¿do not twist or bend the bending section with your hands¿.Additionally, the instructions for safe use manual indicate that, damage to the bending section would happen if excessive force was applied while angulating in the opposite direction if there was no movement from the bending section.
 
Event Description
It was reported that during inspection the insulation test failed.It was noted that the scope has broken and protruded skeleton with excessive broken fibers.No injury to patient or anyone else.
 
Manufacturer Narrative
This supplemental report is submitted to provide the results of the legal manufacturer's investigation and device history record (dhr) review.The dhr for the subject device was reviewed and it was verified the device was manufactured in accordance with documented specifications.Based on similar reported complaints and the device evaluation, the legal manufacturer attributes the likely cause to user handling or technique.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
URETERO-RENO FIBERSCOPE
Type of Device
URETERO-RENO FIBERSCOPE
Manufacturer (Section D)
OLYMPUS MEDICAL SYSTEMS CORP.
2951 ishikawa-cho
hachioji-shi, tokyo-to 192-8 507
JA  192-8507
Manufacturer Contact
kazutaka matsumoto
2951 ishikawa-cho
hachioji-shi, tokyo-to 192-8-507
JA   192-8507
426425177
MDR Report Key10125446
MDR Text Key209399183
Report Number8010047-2020-03159
Device Sequence Number1
Product Code FGB
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K172298
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,user f
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 08/17/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/05/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberURF-P6R
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/07/2020
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Date Manufacturer Received07/21/2020
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-