|
Siemens evaluated the provided information to determined probable root cause.Qc was in range; additional sample draws were tested on the same system to generate the expected <0.006 ng/ml results.The customer used serum samples.Clot time is described by customer as 5 minutes to longer; centrifugation is 3 minutes at 6000 rpm.Allowing samples to clot adequately before centrifugation is important; samples from patients undergoing coagulation therapy may take longer to clot.Fibrin or other particulate matter that is not in the clot and not spun down when using and abbreviated spin time (3 minutes) can cause a falsely elevated result.Preanalytical variables can affect the quality of the sample, and deviation from recommended best practices can lead to erroneous results.While there is insufficient information to determine the cause of the false results, siemens cannot rule out preanalytic factors leading to fibrin interference as the causative agent.Siemens support performed preventative maintenance on the customer instrument; reagent syringe was replaced and a broken acid sensor assembly was replaced.Therefore, siemens also cannot rule out an instrument maintenance influence as part of the potential root cause.The instructions for use states in the summary and explanation of the test states: "always analyze ctni results in the context of time elapsed since patient presentation to the hospital.Serial sampling is recommended to detect the temporal rise and fall of troponin levels characteristic of mi.In accord with published recommendations, serial testing of ctni at intervals of 2 to 4 hours for up to 12 to 24 hours is suggested to corroborate a single ctni result.An elevated troponin alone is not sufficient to make the diagnosis of mi.Other markers, such as ck-mb and myoglobin, can be used in conjunction with ctni results in aiding the diagnosis of mi.A positive troponin result is not always indicative of mi.Other conditions resulting in myocardial cell damage can contribute to elevated cardiac troponin i levels.These conditions include, but are not limited to, myocarditis, cardiac surgery, angina, unstable angina, congestive heart failure, and non-cardiac related causes, such as, renal failure and pulmonary embolism." there are not similar escalations on this tni ultra kit lot and the system is operational.No product performance issue is confirmed.No further evaluation of the device is required.
|