| Model Number |
7753500 |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
05/08/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Neither the device nor films of applicable imaging studies were returned to the manufacturer for evaluation.Therefore, we are unable to determine the definitive cause of the reported event.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
It was reported that the patient presented with t2 spinal adenoma tumor; and underwent posterior cervico-thoracic fusion.Intra-op, the bottom connector set screw broke and stripped when the surgeon was torquing off (tightening) the top connector locking screw.The broken fragments of the bottom set screw got stuck in the top locking screw.The locking screw was explanted completely along with the crosslink.As the bottom set screw had been torqued off (tightened) in the patient, the surgeon did not remove it.No broken fragments of the set screw remained inside the patient.No patient complications were reported as a result of this event.
|
| |
|
Additional Manufacturer Narrative
|
|
H3.Device evaluation summary: visual and optical inspection confirmed the mas connector is broken at the base of the connector hex head.The partial connector is still threaded onto the crosslink.The location of the fracture at the base of the connector head appears to be damaged from torsional overload.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Search Alerts/Recalls
|