AVANOS MEDICAL INC. MICROCUFF ADULT ENDOTRACHEAL TUBE; VAP MICROCUFF ENDOTRACHEAL TUBES Back to Search Results
Model Number 35214
Medical Device Problem Codes Gas/Air Leak (2946); Pressure Problem (3012)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 05/03/2020
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
A review of the device history record is not possible as no lot number was provided.The actual complaint product was not returned for evaluation.Root cause could not be determined.All information reasonably known as of 05 jun 2020 has been included in this health authority report.Should additional information be obtained, a follow-up health authority report will be provided.Avanos medical inc.Has no independent knowledge of the event reported but is relaying the information that was provided by the user facility where the incident occurred.This product incident is documented in the avanos medical inc.Complaint database and identified as complaint (b)(4).
 
Event or Problem Description
Avanos medical inc.Received a single report that referenced four different incidences, which were associated with separate units, involving four different patients.This is the first of four reports.Refer to 3011270181-2020-00084 for the second report.Refer to 3011270181-2020-00085 for the third report.Refer to 3011270181-2020-00086 for the fourth report.It was reported that "several" incidents of subglottic et [endotracheal] tube cuff leaks occurred that are noticed via cuff pressure checks "anywhere from 2-14 days after intubation." anesthesiologist does not always check ett [endotracheal tube] cuff patency prior to intubation.All tubes are placed via video laryngoscope to confirm correct tube placement.When cuff leaks are noticed all patients have required reintubation without difficulty.No patient injuries.Patient information was provided.Date intubated: (b)(6) 2020, date extubated: (b)(6) 2020, peep settings: 6-16, no lot code provided.Additional information received 22-may-2020 indicated the product was not altered in any way prior to use.The problem was discovered through normal patient monitoring.The incident did not occur during surgery, did not occur during transport, and the patient was not hand bagged.It is unknown if the tube migrated.No abnormalities were noted previous to intubation or re-intubation.It is unknown how much pressure was used to inflate the balloon and no patient comorbidity information was provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MICROCUFF ADULT ENDOTRACHEAL TUBE
Common Device Name
VAP MICROCUFF ENDOTRACHEAL TUBES
Manufacturer (Section D)
AVANOS MEDICAL INC.
5405 windward parkway
alpharetta GA 30004
MDR Report Key10126209
Report Number3011270181-2020-00083
Device Sequence Number17046825
Product Code BTR
UDI-Device Identifier10609038352142
UDI-Public10609038352142
Combination Product (Y/N)N
Initial Reporter StateRI
Initial Reporter CountryUS
PMA/510(K) Number
K113333
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 06/05/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number35214
Device Catalogue Number103521401
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/15/2020
Initial Report FDA Received Date06/05/2020
Patient Sequence Number1
Patient Age54 YR
Patient Weight125
-
-