MAUDE Adverse Event Report: GE MEDICAL SYSTEMS (CHINA) CO., LTD (WUXI) CARESTATION 650 A1; ANESTHESIA GAS MACHINE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

GE MEDICAL SYSTEMS (CHINA) CO., LTD (WUXI) CARESTATION 650 A1; ANESTHESIA GAS MACHINE Back to Search Results
Device Problem Loose or Intermittent Connection (1371)
Patient Problem No Patient Involvement (2645)
Event Date 05/12/2020
Event Type  malfunction  
Manufacturer Narrative
No report of patient involvement.The initial reporter is located outside the u.S., and therefore this information is not provided due to country privacy laws.Ge healthcare (gehc) reported a field modification for this issue per 21 cfr 806 on 05 december 2019.The fda recall number is z-0813-2020 & z-0814-2020.Customers were sent a letter explaining the issue and were provided the following safety instructions: you can continue to use the anesthesia system.If you observe the message ventilate manually, change from mechanical to manual ventilation.At any time, the clinician may use a self-inflating bag to ventilate the patient and/or switch to another anesthesia device.Contact your ge healthcare representative for repair of the device.Perform the planned maintenance (pm) every 12-months at a minimum per the user's reference manual which includes inspection of the cable connection.Note: this inspection step is included in the annual pm described in the technical reference manual.Performing this step in the pm would confirm the integrity of the cable connection.
 
Event Description
As a result of an inspection that was completed as part of ge healthcare field modification, this unit was identified as having a loose cable connection that would cause a loss of mechanical ventilation and a high priority audio and visual alarm.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CARESTATION 650 A1
Type of Device
ANESTHESIA GAS MACHINE
Manufacturer (Section D)
GE MEDICAL SYSTEMS (CHINA) CO., LTD (WUXI)
no. 19 changjiang road
national hi-tech dev. zone
wuxi 21402 8
CH  214028
Manufacturer Contact
david woodworth
3030 ohmeda drive
madison, WI 
MDR Report Key10128074
MDR Text Key194315497
Report Number9710602-2020-00176
Device Sequence Number1
Product Code BSZ
Combination Product (y/n)N
PMA/PMN Number
K151570
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,health professional,u
Remedial Action Recall
Type of Report Initial
Report Date 06/08/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Was Device Available for Evaluation? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 03/12/2020
Initial Date FDA Received06/08/2020
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage N
Removal/Correction NumberSEEH10
Patient Sequence Number1
-
-