| Model Number |
RONYX35026UX |
| Medical Device Problem Codes |
Device Damaged by Another Device (2915); Material Deformation (2976)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
05/27/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
A resolute onyx rx coronary, drug eluting stent was used during a procedure to treat a distal svg.It was reported that the stent failed to cross the lesion and when the device was removed the stent was damaged.There was no patient injury reported.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information: a mildly tortuous lesion in the distal circumflex artery.The device was inspected before use with no issues noted.The lesion was pre-dilated.The device did pass through a previously deployed stent.Resistance was noted when advancing the device.It was stated that the resolute onyx device caught on a previously deployed stent.The resolute onyx device was removed and upon inspection the proximal end of the stent had raised struts where it caught on the previously deployed stent.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Search Alerts/Recalls
|