| Model Number |
MT22719 |
| Medical Device Problem Code |
Unable to Obtain Readings (1516)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
06/01/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).
|
| |
|
Event or Problem Description
|
|
It was reported that error icon displayed occurred.No product or data was provided for evaluation.Confirmation of the allegation and a probable cause could not be determined.No injury or medical intervention was reported.
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4).
|
| |
|
Event or Problem Description
|
|
The product was evaluated.An external visual inspection was performed and passed.Receiver charge and boot tests were performed and failed due to the receiver not powering on.Functional tests were not performed due to the receiver not powering on after charging.An internal visual inspection was performed and failed due to moisture damage.Pictures were taken.The receiver log was not downloaded and reviewed due to the receiver not powering on after charging.Confirmation of the allegation and a probable cause could not be determined.No injury or medical intervention was reported.
|
| |
|