COVIDIEN MFG DC BOULDER VALLEYLAB; ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES Back to Search Results
Model Number CTRF100
Medical Device Problem Codes Adverse Event Without Identified Device or Use Problem (2993); Intermittent Energy Output (4025)
Health Effect - Clinical Codes No Code Available (3191); Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available (4581)
Date of Event 05/21/2020
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
(b)(4).If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
According to the reporter, at the middle of ablation procedure, the unit suddenly turned off.Even though the unit was restarted, the indicator turned on but the device did not start up.The procedure was not completed.Reoperation was required due to the issue.The patient was under a local anesthesia when the unit stopped working.The procedure was reschedule the following day and they performed using an alternative device.
 
Additional Manufacturer Narrative
Evaluation summary: medtronic conducted an investigation based upon all information received.The device was available for evaluation.Visual inspection noted the unit had a power supply failure.The resister on the hvdc board was damaged, and there were poor contacts with the power supply.It was reported that the device stopped working unexpectedly during use.A potentially related device issue was confirmed.The most likely cause was traced to a component failure.Internal process improvements have been initiated to mitigate this issue.The manufacturing records for each device are thoroughly reviewed prior to release to ensure that it meets all medtronic quality specifications.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VALLEYLAB
Common Device Name
ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES
Manufacturer (Section D)
COVIDIEN MFG DC BOULDER
5920 longbow dr
boulder CO 80301 3299
MDR Report Key10130187
Report Number1717344-2020-00594
Device Sequence Number2042845
Product Code GEI
UDI-Device Identifier10884524002774
UDI-Public10884524002774
Combination Product (Y/N)N
Initial Reporter CountryJA
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source foreign,user facility
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 12/18/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberCTRF100
Device Catalogue NumberCTRF100
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/28/2020
Initial Date Received by Manufacturer 05/21/2020
Supplement Date Received by Manufacturer09/09/2020
Initial Report FDA Received Date06/08/2020
Supplement Report FDA Received Date12/18/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
-
-