| Medical Device Problem Code |
Difficult to Insert (1316)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
01/20/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).Reported event was confirmed by review of visual evaluation of the returned product.The shell had scratches on the inner radius.The rim has similar dings from a removal tool as were found on the liner.Dimensional analysis not performed as impaction and product damage may interfere with accuracy of readings.Device history record (dhr) was reviewed and no discrepancies were found.Root cause was unable to be determined.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.Multiple mdr reports were filed for this event, please see associated reports: 0001825034-2020-00620.
|
| |
|
Event or Problem Description
|
|
It was reported during surgery, the liner would not assemble with the acetabular shell.This surgery was finished with backup product.No further event information available at the time of this report.
|
| |
|