SMITH & NEPHEW, INC. SUBTROC LAG SCREW HEXDRVR ROD; ROD, FIXATION, INTRMDLLRY AND ACCSSORS Back to Search Results
Model Number 71674086
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 05/12/2020
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that during procedure subtrochanteric driver retaining rod broke at the threading.Smith& nephew backup device was available to complete the procedure.
 
Additional Manufacturer Narrative
The device, used in treatment, was returned for evaluation.A visual inspection confirmed the threads are broken off the subtroc lag screw hexdrvr rod and has not been returned.The device shows significant signs of wear/usage.A review of the manufacturing records did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.At this time, we have no reason to suspect that the products failed to meet any product specifications at the time of manufacture.A complaint history review found related failures; this failure mode will be monitored for future complaints for any necessary corrective actions.Damage from prolonged use, misuse or rough handling are likely probable causes of the reported event.We recommend that all reusable instruments be routinely inspected for wear and damage and replaced as necessary.This is a reusable instrument that can be exposed to numerous surgeries; damage from repeated use can occur.Based on this investigation, the need for corrective action is not indicated.Should additional information be received, the complaint will be reopened.No further investigation warranted for this complaint; however we will continue to monitor for future complaints and investigate as necessary.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SUBTROC LAG SCREW HEXDRVR ROD
Common Device Name
ROD, FIXATION, INTRMDLLRY AND ACCSSORS
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
MDR Report Key10131432
Report Number1020279-2020-02169
Device Sequence Number3342188
Product Code HSB
UDI-Device Identifier03596010580290
UDI-Public03596010580290
Combination Product (Y/N)N
PMA/510(K) Number
K122170
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,consum
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 07/01/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number71674086
Device Catalogue Number71674086
Device Lot Number12FNM0004
Was Device Available for Evaluation? Yes
Initial Date Received by Manufacturer 05/12/2020
Supplement Date Received by Manufacturer05/12/2020
Initial Report FDA Received Date06/08/2020
Supplement Report FDA Received Date07/01/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-