| Model Number |
8725 |
| Medical Device Problem Codes |
Computer Software Problem (1112); Failure to Power Up (1476)
|
| Health Effect - Clinical Code |
No Patient Involvement (2645)
|
| Date of Event |
05/08/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that the system is not booting.A ce motor drive was selected to be used for a procedure.Prior to the procedure, the system was not booting.The procedure was unable to be completed due to this event and it was unknown if the patient was sedated.The patient was stable and no patient complications occurred.
|
| |
|
Event or Problem Description
|
|
It was reported that the system is not booting.A ce motor drive was selected to be used for a procedure.Prior to the procedure, the system was not booting.The procedure was unable to be completed due to this event and it was unknown if the patient was sedated.The patient was stable and no patient complications occurred.
|
| |
|
Additional Manufacturer Narrative
|
|
H6 device codes corrected.The acquisition (acq) and imaging (img) computers (pc) were returned for analysis.A functional test and virus scan of the img pc revealed no issues.A functional test of the acq pc revealed the acq pc failed start up; the power supply was found to be at fault.
|
| |
|