• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY FRANCE SAS - 3003895575 DXTEND MODULAR HUM STEM D10 HA; DELTA XTEND IMPLANTS : SHOULDER HUMERAL STEMS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY FRANCE SAS - 3003895575 DXTEND MODULAR HUM STEM D10 HA; DELTA XTEND IMPLANTS : SHOULDER HUMERAL STEMS Back to Search Results
Model Number 130710000
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Unspecified Infection (1930); Pain (1994); No Code Available (3191)
Date of Event 05/29/2020
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
(b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
Revision shoulder arthroplasty: (b)(6) 2020.Primary implant date: (b)(6) 2017.Painful, infected prosthetic joint, according to surgeon.All prosthesis removed.Patient outcome: stable shoulder joint.No ae reported.No delay in the surgery.Female patient.(b)(6).
 
Additional Manufacturer Narrative
Product complaint # (b)(4).Investigation summary: no device associated with this report was received for examination.A worldwide complaint database search found no other reported incident(s) against the provided product/lot combination(s) since release for distribution.A manufacturing record evaluation was performed for the finished device lot number, and no non-conformances were identified.Based on previous investigations this complication of joint replacement is unlikely to have been the result of a device failing to meet required specifications.The information received will be retained for potential series investigations if triggered by trend analysis, post market surveillance, or other events within the quality system.Corrective action was not indicated.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DXTEND MODULAR HUM STEM D10 HA
Common Device Name
DELTA XTEND IMPLANTS : SHOULDER HUMERAL STEMS
Manufacturer (Section D)
DEPUY FRANCE SAS - 3003895575
7 allee irene joliot-curie
b.p. 256
saint priest cedex 69801
FR  69801
MDR Report Key10133850
Report Number1818910-2020-13448
Device Sequence Number19799968
Product Code KWS
UDI-Device Identifier10603295027126
UDI-Public10603295027126
Combination Product (Y/N)N
Initial Reporter CountryAS
PMA/510(K) Number
K021478
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2017
Device Explanted Year2020
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 05/29/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model Number130710000
Device Catalogue Number130710000
Device Lot Number5285265
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 05/29/2020
Supplement Date Received by Manufacturer07/09/2020
Initial Report FDA Received Date06/09/2020
Supplement Report FDA Received Date07/10/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
DXTEND GLENOSPHERE ECC D38MM; DXTEND METAGLENE; DXTEND MOD EPI 1 ECC LEFT HA; DXTEND MODULAR HUM STEM D10 HA; DXTEND STAND PE CUP D38 +3MM; DXTEND GLENOSPHERE ECC D38MM; DXTEND METAGLENE; DXTEND MOD EPI 1 ECC LEFT HA; DXTEND MODULAR HUM STEM D10 HA; DXTEND STAND PE CUP D38 +3MM
Outcome Attributed to Adverse Event Required Intervention;
Patient Age56 YR
-
-