|
This follow-up report is being submitted to relay additional information. the following sections were updated: b4, b5, g4, g7, h1, h2, h3, h4, h6, h10.One monoblock inserter handle was returned and evaluated.Upon visual inspection the device fractured at the threaded tip.There is no further visible damage to the device.Complaint sample was evaluated and the reported event was confirmed.Device history record was reviewed and no discrepancies were found.Root cause was unable to be determined.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|