• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY MITEK LLC US 5.5 HEALIX PEEK ANCH.W/OCORD; SOFT-TISSUE ANCHOR, BIOABSORBABLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY MITEK LLC US 5.5 HEALIX PEEK ANCH.W/OCORD; SOFT-TISSUE ANCHOR, BIOABSORBABLE Back to Search Results
Model Number 222209
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code Not Applicable (3189)
Date of Event 05/13/2020
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Product complaint # (b)(4).Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.Investigation summary: the complaint device is not being returned, it was discarded by the customer, therefore unavailable for a physical evaluation.With the information provided, and without the complaint device to evaluate, we cannot determine a root cause for the reported failure.A manufacturing record evaluation was performed for the finished device lot number (l680989), and no non-conformances were identified.At this point in time, no corrective action is required, and no further action is warranted.However, depuy synthes mitek will continue to track any related complaints within this device family as a means of monitoring the extent with which this complaint is observed in the field.Patient identifier: (b)(6).Udi: (b)(4).
 
Event or Problem Description
It was reported by sales rep via complaint submission tool that during an unknown procedure, while using 5.5 mm healix peek anchor w/orthocord, the tip broke off before inserting implant into the hole, doctor reached for another implant and it broke again.A mitek fastin anchor was used to complete the surgery with a delay of 5 minutes.No patient consequences reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
5.5 HEALIX PEEK ANCH.W/OCORD
Common Device Name
SOFT-TISSUE ANCHOR, BIOABSORBABLE
Manufacturer (Section D)
DEPUY MITEK LLC US
325 paramount drive
raynham MA 02767
MDR Report Key10135503
Report Number1221934-2020-01523
Device Sequence Number16948954
Product Code MAI
UDI-Device Identifier10886705007578
UDI-Public10886705007578
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
K150209
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 05/26/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date11/30/2020
Device Model Number222209
Device Catalogue Number222209
Device Lot NumberL680989
Was Device Available for Evaluation? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 05/26/2020
Initial Report FDA Received Date06/09/2020
Patient Sequence Number1
-
-