MAUDE Adverse Event Report: DEPUY FRANCE SAS - 3003895575 HO NECK SEGMENT COR AMT; HIP INSTRUMENTS : FEMORAL TRIALS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY FRANCE SAS - 3003895575 HO NECK SEGMENT COR AMT; HIP INSTRUMENTS : FEMORAL TRIALS Back to Search Results
Model Number L20433
Medical Device Problem Codes Device-Device Incompatibility (2919); Material Integrity Problem (2978)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 05/25/2020
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Product complaint # (b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
The surgeons was attempting to trial the high offset neck trial on the size 12 broach as per normal procedure, but was unable to seat the trial neck on the size 12 broach.After being unable to trial the neck, the surgeon then had to implant the definitive stem and then trial off that.Upon investigation by the team after the case, they could see that the adaptor on the broach may be damaged preventing the neck trial from seating.They also tried the std neck trial to confirm, and this was hard to seat on the broach too.No rep was present at the case.
 
Additional Manufacturer Narrative
Product complaint #: (b)(4).Investigation summary: the device was reviewed by depuy engineering melbourne in a-4281685 which states my assessment is that this complaint is due to wear and tear.Root cause attributed to the device being worn from normal use and servicing.Depuy synthes considers the investigation closed at this time.Should additional information be received, the information will be reviewed and the investigation may be re-opened as necessary.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).Investigation summary: no device associated with this report was received for examination.A worldwide lot specific complaint database search, or device history record (dhr) review was not possible because the required lot code(s) was not provided.Based on previous investigations this complication of joint replacement is unlikely to have been the result of a device failing to meet required specifications.The information received will be retained for potential series investigations if triggered by trend analysis, post market surveillance, or other events within the quality system.Corrective action was not indicated.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes joint reconstruction, or its employees that the report constitutes an admission that the product, depuy synthes joint reconstruction, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HO NECK SEGMENT COR AMT
Common Device Name
HIP INSTRUMENTS : FEMORAL TRIALS
Manufacturer (Section D)
DEPUY FRANCE SAS - 3003895575
7 allee irene joliot-curie
b.p. 256
saint priest cedex 69801
FR  69801
MDR Report Key10135642
Report Number1818910-2020-13523
Device Sequence Number19561068
Product Code LXH
UDI-Device Identifier10603295258230
UDI-Public10603295258230
Combination Product (Y/N)N
Initial Reporter CountryNZ
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup,Followup
Report Date (Section B) 05/25/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberL20433
Device Catalogue NumberL20433
Device Lot Number5064961
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 05/25/2020
Supplement Date Received by Manufacturer07/03/2020
07/17/2020
07/23/2020
Initial Report FDA Received Date06/09/2020
Supplement Report FDA Received Date07/06/2020
07/21/2020
07/24/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-