| Model Number |
5319G |
| Medical Device Problem Code |
Patient-Device Incompatibility (2682)
|
| Health Effect - Clinical Codes |
Skin Irritation (2076); Pressure Sore/Ulcer (2326)
|
| Date of Event |
05/22/2020
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
Patient was positioned on mizuho black wilson frame for spine procedure.Procedure lasted 3 hours 13 minutes.After procedure patient flipped back supine.Red, blistered areas noted down both sides of patient where the patient was resting on the foam pads.Pads had been replaced on the frame around six months ago.Frame has been used many times since then.Disposable sheet from our custom turnover kit was used to cover the frame.Silvadene cream used to treat patient wounds.Patient discharged and will receive treatment in outpatient wound center.
|
| |
|
Additional Manufacturer Narrative
|
|
A disposable sheet from a turnover kit was placed on the frame between the patient and the pads.The pads did not show wear and have been in use for several months.The patient was not obese, nor was the case long.The use of extra material between the patient and the pads combined with the patient noted perspiration likely contributed to the pressure issues reported.The customer indicated that since reporting this issue, they have continued to use the frame with no further incidents.
|
| |
|
Event or Problem Description
|
|
Patient was positioned on mizuho black wilson frame for spine procedure.Procedure lasted 3 hours 13 minutes.After procedure patient flipped back supine.Red, blistered areas noted down both sides of patient where the patient was resting on the foam pads.Pads had been replaced on the frame around six months ago.Frame has been used many times since then.Disposable sheet from our custom turnover kit was used to cover the frame.Sivadene cream used to treat patient wounds.Patient discharged and will receive treatment in outpatient wound center.
|
| |
|
Search Alerts/Recalls
|