• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: CONVATEC DOMINICAN REPUBLIC INC ACTIVELIFE; POUCH, COLOSTOMY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

CONVATEC DOMINICAN REPUBLIC INC ACTIVELIFE; POUCH, COLOSTOMY Back to Search Results
Model Number 022765
Medical Device Problem Code Malposition of Device (2616)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Mdr 9618003-2020-11847 / device 1 of 4.Based on the available information, this event is deemed to be a reportable malfunction.To date no additional information has been received.Should additional information become available, a follow-up report will be submitted.(b)(4).
 
Event or Problem Description
The end user reported that the pre-cut opening was off centered in 4 out of 20 wafers from 2 market units and was noted prior to use.The products were not used.No photo is available at this time.
 
Event or Problem Description
To date no additional patient or event details have been received.
 
Additional Manufacturer Narrative
Additional information - this emdr is being submitted to include the below: d4: unique identifier (udi) h6: investigation results under imdrf cause investigation code, imdrf investigation findings, imdrf cause conclusions h10: investigation summary batch record review: the batch record review supports that there were no discrepancies related to the issue reported.This batch was manufactured following the applicable procedures, all machine and testing parameters were in accordance with the applicable procedures and specifications, the testing results were found satisfactory and no nonconformance related to the malfunction code in analysis was noted.The line clearance was executed per standard operating procedure (sop) ¿line clearance procedure and security check¿, the results were documented, and no issues were identified during the manufacturing process.The materials comply with the correct parameters, procedure instruction (pi), ds, mt, tm and the results are satisfactory.Photograph, video and/or physical sample evaluation: no photographs associated with this case were received and no unused return sample was expected.Executive summary of the non-conformance: based on the results of the preliminary investigation, the batch record reviews were performed, and no discrepancies were found.A revision of the improvements made to the process was performed, and actions have been implemented for this type of incidents.In addition, an upgrade of the guard robot arm, disc collar and pouch assembly process was performed.These improvements performed to the guard machine, contribute to the reduction of possible defects related to off center.The last batch received was manufactured prior to the implementation of these improvements.Furthermore, a complaints search conducted shows that no adverse trend was identified from january 2019 to august 2020 for this failure mode.This failure mode will keep monitoring for track and trending.For this reason, no additional investigation is needed.The investigation associated with related event has been approved and is complete.No additional action is required, and this complaint will be closed.This issue will be monitored through the post market product monitoring review process.To date no additional information has been received.Should additional information become available, a follow-up report will be submitted.Fda registration number reporting site: 1049092 manufacturing site: 9618003.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ACTIVELIFE
Common Device Name
POUCH, COLOSTOMY
Manufacturer (Section D)
CONVATEC DOMINICAN REPUBLIC INC
km 18.5 parque industrial
itabo, s.a. haina
san cristobal
Manufacturer (Section G)
CONVATEC DOMINICAN REPUBLIC INC
km 18.5 parque industrial
itabo, s.a. haina
san cristobal
Manufacturer Contact
pamela meadows
7815 national service road
suite 600
greensboro, NC 27409
3365424679
MDR Report Key10136964
Report Number9618003-2020-11847
Device Sequence Number16667615
Product Code EZQ
UDI-Device Identifier00768455179190
UDI-Public00768455179190
Combination Product (Y/N)N
Initial Reporter StateCA
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Consumer
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 05/26/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Expiration Date04/13/2025
Device Model Number022765
Device Lot Number0D01366
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 05/26/2020
Supplement Date Received by Manufacturer11/02/2023
Initial Report FDA Received Date06/10/2020
Supplement Report FDA Received Date11/14/2023
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured04/14/2020
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Patient SexUnknown
-
-