| Model Number |
MMT-751LNAB |
| Medical Device Problem Code |
Device Difficult to Program or Calibrate (1496)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
06/04/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).Currently it is unknown whether or not the device may have caused or contributed to the event as no product has been returned.The device will be returned for analysis and further information will follow once the analysis has been completed.No conclusion can be drawn at this time.
|
| |
|
Event or Problem Description
|
|
Information received by medtronic indicated that from the last few weeks that the buttons were need to be pressed multiple times to work, then it stopped working.Customer reported that the time clock was advancing.The screen.No harm requiring medical intervention was reported.The insulin pump will be returned for analysis.
|
| |
|
Additional Manufacturer Narrative
|
|
Device received with no button response at down due to unlocked j2/lcd keypad connector during visual inspection.Unable to perform operating currents, self test and displacement test due to no button response.No button error alarm during testing.However, keypad flattened button dome switch was found on down.
|
| |
|
Search Alerts/Recalls
|