The initial reporter stated they received discrepant results for two patient samples tested with etoh2 ethanol gen.2 on a cobas 6000 c (501) module.No incorrect results were reported outside of the laboratory.The reporter replaced the ethanol reagent pack and repeated the samples.The reporter also noted there were bubbles in the analyzer cuvettes.The first sample initially resulted with an ethanol value of 30 mg/dl, which repeated as 0.7 mg/dl.The repeat value was believed to be correct.The second sample initially resulted with an ethanol value of 8, which repeated as 0.0.The c501 analyzer serial number is (b)(4).
|