| Model Number |
CD3257-40 |
| Medical Device Problem Codes |
Premature Discharge of Battery (1057); Battery Problem (2885); No Pacing (3268)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
The device is included in the battery performance alert advisory issued by abbott on 28 august 2017.The results of the investigation are inconclusive since the device was not returned for analysis.Based on the information received, the cause of the reported incident could not be conclusively determined.Further information was requested but not received.
|
| |
|
Event or Problem Description
|
|
Following the battery performance alert (bpa) advisory, a bpa was received by the clinician and the device was explanted and replaced.It was noted that the implantable cardioverter defibrillator failed to pace and it was alleged that the device exhibited premature battery depletion.
|
| |
|
Additional Manufacturer Narrative
|
|
Analysis determined that the implantable cardioverter defibrillator was in bvvi due to a power on reset, which then resulted in the device no longer pacing in the atrial chamber.The analysis confirmed that the por was likely caused by a prematurely depleted battery.No sources of high current were noted.The cause of the premature battery depletion was consistent with lithium (li) cluster formation.From these analyses, in the absence of high current draw, it is probable that the premature battery depletion was caused by a lithium cluster induced short circuit.Li clusters are a known depletion mechanism for these advisory products that has been investigated and associated with a field action in october 2016.
|
| |
|