• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITHS MEDICAL ASD, INC. CADD ADMINISTRATION SET; SET, ADMINISTRATION, INTRAVASCULAR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITHS MEDICAL ASD, INC. CADD ADMINISTRATION SET; SET, ADMINISTRATION, INTRAVASCULAR Back to Search Results
Model Number 21-7381-24
Medical Device Problem Codes Fluid/Blood Leak (1250); Material Fragmentation (1261)
Health Effect - Clinical Code No Information (3190)
Date of Event 05/15/2020
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(6).Not returned: (b)(6) 2020; 02/15/2025.Occupation: regional nurse manager.
 
Event or Problem Description
Information was received indicating that a smiths medical cadd administration set tubing got detached during the administration of total parenteral nutrition (tpn).The end of the administration set fell off.There was no reported adverse event.
 
Additional Manufacturer Narrative
Device evaluation: three pictures of cadd administration sets were received by shm for analysis, and in two of them its valve detached from the tube was observed and the third picture showed a tpn product coiled.No testing could be performed because no samples were provided to perform a thorough investigation.The customer's reported problem was confirmed.According to the investigation, the most probable root cause is that the use of dispenser solvent in without stopper originated a lack of solvent.According to the investigation production personnel was trained in order to reinforce the solvent application.The problem source of the reported problem is manufacturing.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CADD ADMINISTRATION SET
Common Device Name
SET, ADMINISTRATION, INTRAVASCULAR
Manufacturer (Section D)
SMITHS MEDICAL ASD, INC.
6000 nathan lane north
minneapolis MN 55442
MDR Report Key10139102
Report Number3012307300-2020-05735
Device Sequence Number2309464
Product Code FPA
UDI-Device Identifier10610586027468
UDI-Public10610586027468
Combination Product (Y/N)N
Initial Reporter StateVA
Initial Reporter CountryUS
PMA/510(K) Number
K031361
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source consumer,user facility
Initial Reporter Occupation Nurse
Type of Report Initial,Followup
Report Date (Section B) 07/13/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this a Product Problem Report? Yes
Operator of Device No Information
Device Expiration Date02/15/2025
Device Model Number21-7381-24
Device Catalogue Number21-7381-24
Device Lot Number3935900
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Initial Date Received by Manufacturer 05/15/2020
Supplement Date Received by Manufacturer06/18/2020
Initial Report FDA Received Date06/10/2020
Supplement Report FDA Received Date07/13/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age73 YR
Patient Weight50
-
-