MAUDE Adverse Event Report: SMITH & NEPHEW, INC. REF XLPE 32 20 DEG 58-60 G; PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. REF XLPE 32 20 DEG 58-60 G; PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED Back to Search Results
Model Number 71333336
Medical Device Problem Code Activation, Positioning or Separation Problem (2906)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 05/11/2020
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that during a thr procedure, after installing the reflection xlpe 58mm-60mm liner during the surgery, it was found that the liner could not be locked with the outer cup.After trying for 20 minutes, another liner of the same size was opened, and finally the surgery was completed successfully.The event resulted in a total delay of nearly 30 minutes, which increased the risk of bleeding in surgery, nevertheless, no injury to the patient was reported.
 
Additional Manufacturer Narrative
The associated device, used in treatment, was returned and evaluated.A visual inspection of the returned device could not confirm the stated failure mode.The device has signs of damage from attempted use.The device was manufactured in 2019.A dimensional inspection was attempted but the device was too damaged from the attempted insertion to obtain accurate measurements.A review of complaint history did not reveal additional complaints for the listed batch.A review of the manufacturing records did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.At this time, we do not have reason to suspect that the product failed to meet any product specifications at the time of manufacture.The potential probable cause for this event is likely a procedural error.Based on this investigation, the need for corrective action is not indicated.Should additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however, we will continue to monitor for future complaints and investigate as necessary.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
REF XLPE 32 20 DEG 58-60 G
Common Device Name
PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
MDR Report Key10139547
Report Number1020279-2020-02302
Device Sequence Number13997497
Product Code LPH
UDI-Device Identifier03596010455697
UDI-Public03596010455697
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
K002747
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 11/16/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number71333336
Device Catalogue Number71333336
Device Lot Number19HM06348
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/08/2020
Initial Date Received by Manufacturer 05/11/2020
Supplement Date Received by Manufacturer11/11/2020
Initial Report FDA Received Date06/10/2020
Supplement Report FDA Received Date11/16/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-