Model Number ASHA4730-01
Medical Device Problem Code Unexpected Shutdown (4019)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 05/19/2020
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(6).
 
Event or Problem Description
It was reported that, during a hip surgery, the "ambient hipvac 50° ifs wand" stopped working and creating plasma while debriding the labrum.The procedure was successfully completed without significant delay using a back-up device.No patient injuries or other complications were reported.All available information has been disclosed.If additional information should become available, a supplemental report will be submitted accordingly.
 
Additional Manufacturer Narrative
H10: h2, h6.The device, used in treatment, was not returned for evaluation.A relationship, if any, between the subject device and the reported event could not be determined since the product was not returned for evaluation.A review of the device history records showed there were no indications to suggest that the product did not meet manufacturing specification or would not be able to perform as intended.A complaint history review found a related failure; this failure mode will be trended to assess for any necessary corrective actions visual inspection and functional testing could not be performed since the device was not returned for evaluation.Thus, the complaint could not be verified, nor could a root cause be determined with confidence.If the product associated with this event is returned at a future date, this evaluation will be reopened for investigation.Potential factors unrelated to the manufacture or design of the device that could have contributed to the reported event includes: (1) previous use (2) disconnecting the wand from the controller after power has been turned on.(3) an issue with one of the concomitant devices in use at the time of this complaint event.There are no indications to suggest the device did not meet product specifications upon release into distribution.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AMBIENT HIPVAC 50 IFS
Common Device Name
ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES
Manufacturer (Section D)
ARTHROCARE CORP.
7000 w. william cannon
austin TX 78735
MDR Report Key10139862
Report Number3006524618-2020-00360
Device Sequence Number19272972
Product Code GEI
UDI-Device Identifier00817470000436
UDI-Public00817470000436
Combination Product (Y/N)N
Initial Reporter CountrySF
PMA/510(K) Number
K083306
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 07/14/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date09/16/2022
Device Model NumberASHA4730-01
Device Catalogue NumberASHA4730-01
Device Lot Number2041693
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 05/19/2020
Supplement Date Received by Manufacturer05/19/2020
Initial Report FDA Received Date06/10/2020
Supplement Report FDA Received Date07/14/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-