Model Number 7010
Medical Device Problem Code Material Deformation (2976)
Health Effect - Clinical Code No Patient Involvement (2645)
Date of Event 02/28/2020
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Age at time of event: 18 years or older.Device evaluated by mfr.: returned product consisted of a quantum maverick balloon catheter.The shaft, hypotube, tip and balloon were microscopically and visually examined.There were numerous kinks.There was a complete separation at 47.8cm distal of the strain relief.The balloon was tightly folded.Inspection of the remainder of the device presented no other damage or irregularities.Product analysis confirmed the reported kink.
 
Event or Problem Description
Reportable based on device analysis completed on 16 may 2020.It was reported that shaft kink occurred.A 2.5mm x 12mm quantum maverick balloon catheter was selected for use.However, during unpacking, it was noted that the shaft was kinked.However, returned device analysis revealed shaft detached/ separated.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
QUANTUM MAVERICK
Common Device Name
CATHETERS, TRANSLUMINAL CORONARY ANGIOPLASTY, PERCUTANEOUS
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
two scimed place
maple grove MN 55311
MDR Report Key10140285
Report Number2134265-2020-07835
Device Sequence Number2101548
Product Code LOX
UDI-Device Identifier08714729392651
UDI-Public08714729392651
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
P860019
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,distri
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 06/10/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date05/20/2022
Device Model Number7010
Device Catalogue Number7010
Device Lot Number0023825479
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/24/2020
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 05/16/2020
Initial Report FDA Received Date06/10/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-