| Model Number |
7010 |
| Medical Device Problem Code |
Material Deformation (2976)
|
| Health Effect - Clinical Code |
No Patient Involvement (2645)
|
| Date of Event |
02/28/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Age at time of event: 18 years or older.Device evaluated by mfr.: returned product consisted of a quantum maverick balloon catheter.The shaft, hypotube, tip and balloon were microscopically and visually examined.There were numerous kinks.There was a complete separation at 47.8cm distal of the strain relief.The balloon was tightly folded.Inspection of the remainder of the device presented no other damage or irregularities.Product analysis confirmed the reported kink.
|
| |
|
Event or Problem Description
|
|
Reportable based on device analysis completed on 16 may 2020.It was reported that shaft kink occurred.A 2.5mm x 12mm quantum maverick balloon catheter was selected for use.However, during unpacking, it was noted that the shaft was kinked.However, returned device analysis revealed shaft detached/ separated.
|
| |
|