Model Number 1650DE
Medical Device Problem Code Battery Problem (2885)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 04/28/2020
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
This device is used for treatment not diagnosis.The ventricular assist system is indicated for use as a bridge to cardiac transplantation and destination therapy in patients who are at risk of death from refractory end-stage left ventricular heart failure.The system is designed for in-hospital and out-of-hospital settings.Concomitant medical products: 5076-52 lead, 694765 lead, implanted: (b)(6) 2007.Investigation of this event is pending and a supplemental report will be sent upon its completion.If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
It was reported that the battery was giving an erroneous critical battery alarm.The battery had a charge capacity of 75% at the time of the alarm.The battery was exchanged.No patient complications have been reported as a result of this event.
 
Additional Manufacturer Narrative
A supplemental report is being submitted for device evaluation.Product event summary: the battery was returned for evaluation.Failure analysis of the returned battery revealed that the device passed visual inspection and functional testing.Log file analysis revealed that the controller in use during the reported event contained a software with a feature that records whether a power source experienced a communication error or a disconnection within each 15-minute interval.Analysis of the alarm log file revealed a critical battery alarm due to a communication error involving the battery.As a result, the reported event was confirmed.A power source lubrication procedure had been performed on the battery in accordance with the requirements under fsca cvg-18-q4-19 on (b)(6) 2018.The most likely root cause of the reported event can be attributed to a communication error between the controller and battery, which may be due to a momentary disconnection on the communication pins of the controller, the controller not receiving responses from the batteries, and/or due to the packet error checking method detecting bit errors.Investigation of this event is completed and the file will be closed.If new information is received, the file will be re-opened and a supplemental will be submitted.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HEARTWARE VENTRICULAR ASSIST SYSTEM - BATTERY
Common Device Name
VENTRICULAR (ASSIST) BYPASS
Manufacturer (Section D)
HEARTWARE, INC.
14400 nw 60th ave
miami lakes FL 33014
MDR Report Key10140778
Report Number3007042319-2020-03497
Device Sequence Number17053008
Product Code DSQ
UDI-Device Identifier00888707002646
UDI-Public00888707002646
Combination Product (Y/N)N
Initial Reporter StateKY
Initial Reporter CountryUS
PMA/510(K) Number
P100047
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Other Health Care Professional
Remedial Action Recall
Type of Report Initial,Followup
Report Date (Section B) 08/04/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Expiration Date08/31/2019
Device Model Number1650DE
Device Catalogue Number1650DE
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/27/2020
Initial Date Received by Manufacturer 06/03/2020
Supplement Date Received by Manufacturer07/29/2020
Initial Report FDA Received Date06/10/2020
Supplement Report FDA Received Date08/04/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
Z-1903-2018
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
1103 VAD
Patient Age44 YR
-
-