• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: GYRUS ACMI, INC SLIMLINE URETEROSCOPE; URETERO-RENO VIDEOSCOPE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

GYRUS ACMI, INC SLIMLINE URETEROSCOPE; URETERO-RENO VIDEOSCOPE Back to Search Results
Model Number MRO-742A
Device Problem Break (1069)
Patient Problem No Patient Involvement (2645)
Event Type  malfunction  
Manufacturer Narrative
A visual inspection on the device found the shaft broken off from the main body of the device at the proximal end.The metal section on the broken end of the shaft is twisted, which would suggest that excessive force was used to cause the shaft to become detached/broken.The instruction manual states the following; ¿do not twist or bend the bending section with your hands.Equipment damage may result¿.The instructions for safe use manual also indicates that, damage to the bending section would happen if excessive force was applied while angulating in the opposite direction if there was no movement from the bending section; this would occur more so in a narrow environment.Additionally, the image seen through the eyepiece lens is very unclear, which would also likely be attributed from the damaged shaft.Based on the evaluation, the likely cause of the damaged and broken shaft is due to mishandling in the form of excessive force.
 
Event Description
It was reported that the scope has separated from the handle from the actual scope.A part of the scope is broken.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SLIMLINE URETEROSCOPE
Type of Device
URETERO-RENO VIDEOSCOPE
Manufacturer (Section D)
GYRUS ACMI, INC
136 turnpike road
southborough MA 01772
Manufacturer Contact
kenneth pittman
93 north pleasant st.
norwalk, OH 44857
9013785969
MDR Report Key10140865
MDR Text Key217367361
Report Number1519132-2020-00015
Device Sequence Number1
Product Code FGB
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K011849
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,user f
Reporter Occupation Other
Type of Report Initial
Report Date 06/10/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/10/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberMRO-742A
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/18/2020
Was the Report Sent to FDA? No
Date Manufacturer Received05/12/2020
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-