• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ETHICON ENDO-SURGERY, LLC. ENDOPATH XCEL BLUNT TIP TROCAR; LAPAROSCOPE, GENERAL AND PLASTIC SURGERY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ETHICON ENDO-SURGERY, LLC. ENDOPATH XCEL BLUNT TIP TROCAR; LAPAROSCOPE, GENERAL AND PLASTIC SURGERY Back to Search Results
Model Number H12LP
Medical Device Problem Codes Gas/Air Leak (2946); Material Integrity Problem (2978)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 02/26/2020
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).Batch #: unk.A manufacturing record evaluation was performed for the finished device lot number and no non-conformance's were identified.Attempts are being made to retrieve the device.To date, the device has not been returned.If the device or further details are received at a later date a supplemental medwatch will be sent.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Event or Problem Description
Hassan trocar: losing co2, seal torn, opened new device and continued procedure was surgery delayed due to the reported event? unknown, was procedure successfully completed? unknown, were fragments generated? unknown, if yes, were they removed easily without additional intervention? unknown, patient status/ outcome / consequences: no, was other medical intervention (e.G.X-rays, additional procedures, prescriptions, otc, revision) required: unknown, is the patient part of a clinical study: unknown, (b)(4).Device property of: none, device in possession of: none by checking this box i certify that all information that are known/available has been disclosed.If any new information will be made available, the additional information will be submitted through cst.True.
 
Additional Manufacturer Narrative
(b)(4).Date sent: 8/27/2020.D4: batch # u93068.Investigation summary: the analysis results found that the h12lp device was returned with no damage in the external components.A leak test was performed and the device was noted to leak without the test probe inserted through the device.Upon further analysis, the universal seal was noted to be damaged.No conclusion could be reached as to what might have caused the damaged on the device.It should be noted that as part of our quality process, each device is visually inspected and functionally tested during manufacturing to ensure the device meets the required specifications prior to shipment.A manufacturing record evaluation was performed for the finished device batch number, and no non-conformances were identified.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ENDOPATH XCEL BLUNT TIP TROCAR
Common Device Name
LAPAROSCOPE, GENERAL AND PLASTIC SURGERY
Manufacturer (Section D)
ETHICON ENDO-SURGERY, LLC.
475 calle c
guaynabo 00969
MDR Report Key10141072
Report Number3005075853-2020-02976
Device Sequence Number2318466
Product Code GCJ
UDI-Device Identifier10705036002017
UDI-Public10705036002017
Combination Product (Y/N)N
PMA/510(K) Number
K032676
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source foreign,health professional
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 05/15/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date12/07/2024
Device Model NumberH12LP
Device Catalogue NumberH12LP
Device Lot NumberU4008R
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/03/2020
Initial Date Received by Manufacturer 05/15/2020
Supplement Date Received by Manufacturer08/03/2020
Initial Report FDA Received Date06/10/2020
Supplement Report FDA Received Date08/27/2020
Patient Sequence Number1
-
-