The user facility reported that the involved navicross was used during the procedure.The catheter was opened and was being advanced over a glidewire advantage.035 when it was noticed that it was kinked so badly that it would not go over the wire; it was almost severed.It was straight out of the box.A different catheter was selected, and the case proceeded and the procedure was completed.There was no patient injury, medical/surgical intervention required.The patient was in stable condition.
|
This report is being submitted as follow up no.1 to provide the device return date in section d10, update section h3, and to provide the completed investigation results.The actual sample was received for evaluation.Visual inspection revealed that the actual sample had been fracture at approximately 968mm from the distal end and separated in two fragments.They were called section a and b in this report.The total length of section a and b was confirmed to be 1358mm.Compared to that of a current sample of the involved product code, which was 1350mm, it was found that the actual sample had been elongated by approximately 8mm.Section a was inspected under ccd while being rotated by 90° each time, and revealed that the fracture end suggested that the shaft had been pulled and torn off, reinforcing wire was seen on the fracture surface, the fracture end had been flattened.Section b was inspected under ccd while being rotated by 90° each time, and revealed that the fracture end suggested that the shaft had been pulled and torn off, reinforcing wire was seen on the fracture surface, the fracture end had been flattened and scratched.Based on these finding, it is likely that the actual sample was trapped by some hard object.The fracture end of section a and b seen from the front confirmed the reinforcing wire had been exposed and the outer layer had been stretched.It is likely that the actual sample in a state of having been trapped with a hard object might have been subjected to excessive pulling force.Magnifying inspection of section b on the part other than the fracture end revealed that the shaft had been kinked at approximately 30mm from the distal end, a bend was found at approximately 390mm from the distal end.The outer and inner diameters of the actual sample were measured on the undamaged part and confirmed to meet the manufacturer control criteria.Ifu states: carefully handle the product under fluoroscopy.If any resistance is felt while handing the product, immediately stop the manipulation and find out cause of the resistance in order to avoid damage to blood vessels and separation or breakage of the product.Based on the provided information and investigation results, there is no definitive evidence that this event was related to a device defect or malfunction.It is likely that the actual sample may have been trapped due some factors, and then subjected to excessive pulling force, resulting in the generation of the fracture.However, the exact cause of the reported event cannot be definitively determined based on the available information.
|