• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ENCORE MEDICAL L.P. RSP; SOCKET, INSERT 36MM SEMI RSP HUMERAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ENCORE MEDICAL L.P. RSP; SOCKET, INSERT 36MM SEMI RSP HUMERAL Back to Search Results
Model Number 508-02-044
Health Effect - Clinical Code No Code Available (3191)
Date of Event 05/14/2020
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Additional reporting on this event will be provided as a supplemental report to these documents as soon as it becomes available.
 
Event or Problem Description
Second revision surgery - dislocated after revision surgery.
 
Additional Manufacturer Narrative
Corrected data: see d.1., d.2., d.4., d.6., e.1.& g.5.Manufacturer narrative: the reason for this revision surgery was reported as dislocation.The previous surgery and the surgery detailed in this event occurred 3 years apart.Initial or prolonged hospitalization was required.The healthcare professional indicated there was no delay in surgery and another suitable device was available for use.The revision surgery was completed as intended.The device was disposed of at hospital and not made available to djo surgical for examination.A review of the device history record (dhr) show that the reported component used in the previous surgery, when released for use, met design and manufacturing requirements.There were no non-conforming material reports (ncmr) associated with the product that may have contributed to the reported event.The device was verified to have gone through an acceptable sterilization process and was within its expiration date at the time of the previous surgery.Customer complaint history of the reported device showed no present trends or on-going issues that are needing a review.The root cause of this complaint was a revision surgery due to dislocation.There were no findings during this evaluation that indicate the reported device was defective.No information was submitted with the complaint regarding pre-existing conditions of the patient or any activities the patient was involved in that may have contributed to the event.There are multiple factors that may contribute to an event that are outside the control of djo surgical.There are no indications of a product or process issue affecting implant safety or effectiveness.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RSP
Common Device Name
SOCKET, INSERT 36MM SEMI RSP HUMERAL
Manufacturer (Section D)
ENCORE MEDICAL L.P.
9800 metric blvd
austin, tx 78758-5445,
MDR Report Key10143264
Report Number1644408-2020-00441
Device Sequence Number19698804
Product Code PHX
UDI-Device Identifier00190446257028
UDI-Public(01)00190446257028
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
K051075
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2017
Device Explanted Year2020
Reporter Type Manufacturer
Report Source company representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 07/13/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date01/27/2022
Device Model Number508-02-044
Device Catalogue Number508-01-036
Device Lot Number860C1913
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 05/14/2020
Supplement Date Received by Manufacturer06/15/2020
Initial Report FDA Received Date06/11/2020
Supplement Report FDA Received Date07/13/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
Patient Age65 YR
-
-