Model Number CI-1500-04
Device Problem Electro-Static Discharge (2149)
Patient Problem Pain (1994)
Event Date 07/13/2020
Event Type  malfunction  
Event Description
The recipient reportedly experienced an electrostatic discharge event.The recipient is presenting with pain.Programming adjustments were made, however, the issue did not resolve.The recipient was advised to discontinue use of device.Revision surgery will be scheduled.
 
Manufacturer Narrative
The recipient's device was reportedly explanted.The recipient was reimplanted with another advanced bionics cochlear device.Disclaimer: advanced bionics does not intend that this report be any admission of liability, fault or product defect.
 
Manufacturer Narrative
Advanced bionics considers the investigation into this reportable event as closed.The external visual inspection revealed the electrode was severed prior to receipt.This is believed to have occurred during revision surgery.The device passed the photographic imaging inspection.System lock was verified.The electrode condition prevented some electrical tests from being performed.The device passed some of the electrical tests performed.The device passed the mechanical tests performed.The failure of this device is attributed to a short from the power node to the case ground at the analog chip.It is believed that electrostatic discharge (esd) led to an overload voltage, damaging structures on the case ground node inside the analog integrated circuit.This ultimately caused the device to cease functioning.A corrective action was implemented.This is the final report.Disclaimer: advanced bionics does not intend that this report be any admission of liability, fault or product defect.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HIRES 90K ADVANTAGE IMPLANT
Type of Device
COCHLEAR IMPLANT
Manufacturer (Section D)
ADVANCED BIONICS, LLC
28515 westinghouse place
valencia CA 91355
MDR Report Key10144688
MDR Text Key195233210
Report Number3006556115-2020-00444
Device Sequence Number1
Product Code MCM
UDI-Device Identifier07630016828638
UDI-Public(01)07630016828638(11)150915(17)180731
Combination Product (y/n)N
PMA/PMN Number
P960058
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Type of Report Initial,Followup,Followup
Report Date 05/21/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/11/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date07/31/2018
Device Model NumberCI-1500-04
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/15/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Age62 YR
-
-