| Model Number |
CD2211-36Q |
| Medical Device Problem Code |
Defective Alarm (1014)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
05/27/2020
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Further information was requested but not received.The results of the investigation are inconclusive since the device was not returned for analysis.Based on the information received, the cause of the reported incident could not be conclusively determined.
|
| |
|
Event or Problem Description
|
|
During a clinic follow-up, a vibratory alert was unable to be delivered by the device during testing.Technical support was contacted, but no intervention has been performed at this time.The patient was stable and will continue to be monitored.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information: h3 and h6.The reported field event of patient notifier anomaly was confirmed in the lab.Interrogation of the device revealed that the device was above the elective replacement indicator (eri) when received for analysis.A longevity calculation was performed based upon programmed settings and usage data in the device image.The calculations revealed that the battery depletion was normal up to the event date.The device's functionality was bench tested regarding telemetry, pacing, sensing, impedance, high voltage (hv) output, and hv shock.The patient notifier failed to vibrate during the bench test.The cause of the anomaly was due to a patient notifier component failure, however, the cause of the failed patient notifier component could not be conclusively undetermined.
|
| |
|
Additional Manufacturer Narrative
|
|
Further information was requested but not received.
|
| |
|
Event or Problem Description
|
|
Additional information received that the device was explanted to resolve the event.The patient was stable and will continue to be monitored.
|
| |
|