A & I INDUSTRIES LTD DRIVE NITRO; ROLLATOR Back to Search Results
Model Number 10266CF
Medical Device Problem Code Material Fragmentation (1261)
Health Effect - Clinical Code Fall (1848)
Date of Event 06/18/2018
Type of Reportable Event Serious Injury
Event or Problem Description
Drive devilbiss healthcare is the initial importer of the device which is a rollator.We are filing this report in an overabundance of caution sue to a mdr regression analysis.The end-user was on the steps of a swimming pool while sitting on the device.The right front wheel reportedly fell off causing him to hit his head.He received minor abrasions as well.The device was received for evaluation.It showed normal usage.The brakes on each side were conforming.The front left wheel was detached from the unit.The front axle bolt was delivered with no damaged and signs of stripped threads.The threads inside the fork, however, show minimal damage to the threads.The root cause may be attributed to these damage threads from assembly, causing the axle bolt to detach from the unit during vibrations over time.The unit is defective historical data for complaints of the same nature was reviewed and found very limited reporting which shows no trend.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DRIVE NITRO
Common Device Name
ROLLATOR
Manufacturer (Section D)
A & I INDUSTRIES LTD
lian du industry park
le liu town, shunde district
foshan city, guangdong
MDR Report Key10146400
Report Number2438477-2018-00068
Device Sequence Number16875000
Product Code ITJ
UDI-Device Identifier00822383288819
UDI-Public822383288819
Combination Product (Y/N)N
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Importer
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 06/10/2020
Report Date (Section F) 06/10/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Lay User/Patient
Device Model Number10266CF
Was Device Available for Evaluation? Yes
Was the Report Sent to FDA? Yes
Distributor Facility Aware Date06/18/2018
Device Age16 MO
Event Location Other
Date Report to Manufacturer06/12/2020
Initial Date Received by Manufacturer Not provided
Initial Report FDA Received Date06/11/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient Weight101
Date Report Sent to FDA06/10/2020
-
-