| Model Number |
401914 |
| Medical Device Problem Code |
Use of Device Problem (1670)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
03/25/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The device was returned to stryker sustainability solutions for evaluation.Visual inspection revealed a bend 4.5 cm from the distal tip.The device did not meet curve and planarity specification.A review of the dhr supports that the device met all inspection and test criteria prior to release from stryker.The most likely root cause is mishandling subsequent to distribution, including shipping/storage conditions or improper manipulation of the catheter.The instructions for use (ifu) state: inspect the catheter for overall condition and physical integrity.Do not use the catheter if electrodes appear loose or if any damage is noted.If such problems exist, return the catheter and packaging to stryker sustainability solutions.Connect the catheter to the corresponding cable connector.Connect the cable to the correct electronic equipment for recording and/or sensing.Observe polarity of proximally located connector pins of the interface cable when connecting to the electronic equipment.Isolate any unused connector pins to reduce development of accidental current pathways to the heart.Follow a suitable electrophysiology study protocol.Do not autoclave catheter.Do not use for electrical ablation.Avoid excessive contact of handpiece with fluids, as this could adversely affect the electrical performance of the catheter.Diagnostic ep catheters are not recommended for long term pacing.The reported event will continue to be monitored through post-market surveillance.
|
| |
|
Event or Problem Description
|
|
It was reported the catheter was not working properly.Multiple attempts were made for additional information from the facility, however none were provided.There was no patient injury, medical intervention, or extended procedure time reported.These are commonly used devices that are readily available.
|
| |
|
Search Alerts/Recalls
|