• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION AMS INFLATABLE PENILE PROSTHESIS; DEVICE IMPOTENCE MECHANICAL/HYDRAULIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION AMS INFLATABLE PENILE PROSTHESIS; DEVICE IMPOTENCE MECHANICAL/HYDRAULIC Back to Search Results
Model Number UNK-P-IPP
Device Problems Collapse (1099); Failure to Cycle (1142); Inflation Problem (1310); Defective Device (2588)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 05/27/2020
Event Type  malfunction  
Event Description
It was reported that during a phone call with the physician it was indicated that he had pump activation issues with three different inflatable penile prosthesis (ipp) devices.It was specified that the pump would "lock up", felt "rock hard" and was not able to be reset.The pumps were said to be so hard that they would stay collapsed.When squeezed they would not refill and after resetting the component it would temporarily work but then collapse after the following initial squeeze.The physician speculated there was a small leak as overtime the pumps would refill.Two of the patients are currently awaiting a future revision.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AMS INFLATABLE PENILE PROSTHESIS
Type of Device
DEVICE IMPOTENCE MECHANICAL/HYDRAULIC
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
10700 bren road w
minnetonka MN 55343
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
10700 bren road w
minnetonka MN 55343
Manufacturer Contact
alyson harris
10700 bren road w
minnetonka, MN 55343
4083953452
MDR Report Key10146535
MDR Text Key194935514
Report Number2183959-2020-02495
Device Sequence Number1
Product Code FHW
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Reporter Occupation Physician
Type of Report Initial
Report Date 06/11/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Model NumberUNK-P-IPP
Device Catalogue NumberUNK-P-IPP
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 05/27/2020
Initial Date FDA Received06/11/2020
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Type of Device Usage Initial
Patient Sequence Number1
-
-