| Model Number |
1219-28-256 |
| Medical Device Problem Codes |
Naturally Worn (2988); Noise, Audible (3273)
|
| Health Effect - Clinical Codes |
Adhesion(s) (1695); Pain (1994); No Code Available (3191)
|
| Date of Event |
05/06/2020
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
(b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Event or Problem Description
|
|
Surgeon reported to bfarm: "increasing pain in the right hip without trauma after getting up and audible cracking; preoperative radiological: ceramic head fracture after hip tep 15 years ago; intraoperatively: the most likely older event with fragments that are firmly encapsulated in the capsule / scar tissue, some freshly lying in the joint.".
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).Investigation summary: examination of the returned liner finds nothing outward to suggest product error or any contribution to the reason for revision.The information received will be retained for potential series investigations if triggered by trend analysis, post market surveillance, or other events within the quality system.Depuy considers the investigation closed.Should additional information be received, the information will be reviewed, and the investigation will be re-opened as necessary.
|
| |
|