Model Number CV-180 |
Device Problem
Power Problem (3010)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Date 05/18/2020 |
Event Type
malfunction
|
Manufacturer Narrative
|
The subject device has not returned to olympus medical systems corp.(omsc) for evaluation.There were no further details provided.If significant additional information is received, this report will be supplemented.
|
|
Event Description
|
During the inspection of use the subject device, the subject device could not be turned the power on.Olympus service operation repair center replaced the power supply unit of the subject device to the new power supply unit.Other detailed information was not provided.There was no report of patient injury associated with the event.
|
|
Manufacturer Narrative
|
This is a supplemental report for mfr report.Olympus medical systems corp.(omsc) could not investigate the subject device, because the subject device was not returned to omsc.Since the subject device was not returned, the exact cause was unknown.The instruction manual of the subject device states the corresponding method in case of an abnormality.
|
|
Manufacturer Narrative
|
This report is being supplemented to provide correction and additional information based on the legal manufacturer's final investigation.A review of the device history record found no deviations that could have caused or contributed to the reported issue.Based on the results of the investigation, and since six years or more have passed since the delivery of the device, it is likely the parts of the power supply unit deteriorated and the power supply unit failed due to repeated use for a long period of time.
|
|