| Model Number |
06870287001 |
| Medical Device Problem Codes |
Fire (1245); Melted (1385)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
05/17/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported device exploded and caught fire and was melted.The patient stated she heard a loud "pop" and saw black smoke and flames coming from her meter carrying case and she doused the meter with water.
|
| |
|
Additional Manufacturer Narrative
|
|
As apparent during the examination of the returned meter, most of the damage to the meter was to the front area by the strip port.The aviva ii meter uses a 3v battery.This battery type does not have enough voltage/capacity to ignite/explode by itself with an external factor.Per the meter labeling, the meter should not be stored by a heat source.Due to the condition of the returned meter it has been concluded that the damage did not originate from the meter itself and has been determined to be a result of customer mishandling.
|
| |
|
Additional Manufacturer Narrative
|
|
This supplemental mdr is being submitted as a part of a retrospective review and remediation effort based on errors that occurred in the submission of mdrs for incorrect manufacturer name and/or address.A non-conformance report (ncr) ¿ 16847 to implement corrective actions was opened on january 29, 2025.Device performance and/or risk profile are not impacted.
|
| |
|