• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ALLERGAN (COSTA RICA) STYLE 10 SILICONE GEL FILLED BREAST IMPLANT; PROSTHESIS, BREAST, NONINFLATABLE, INTERNAL, SILICONE GEL-FILLED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ALLERGAN (COSTA RICA) STYLE 10 SILICONE GEL FILLED BREAST IMPLANT; PROSTHESIS, BREAST, NONINFLATABLE, INTERNAL, SILICONE GEL-FILLED Back to Search Results
Model Number 10-450
Medical Device Problem Codes Gel Leak (1267); Patient-Device Incompatibility (2682)
Health Effect - Clinical Code Capsular Contracture (1761)
Date of Event 05/14/2020
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
A review of the device history record has been completed.No deviations or non-conformances noted.The event of capsular contracture is a physiological complication and analysis of the device generally does not assist allergan in determining a probable cause for this event.The reason for reoperation: capsular contracture baker grade iii.Further information from the reporter regarding event, product, or patient details has been requested.No additional information is available at this time.
 
Event or Problem Description
Healthcare provider reported left side capsular contracture baker grade iii.Device remains implanted.
 
Event or Problem Description
Healthcare provider additionally reported "gel bleed." device has been explanted.
 
Additional Manufacturer Narrative
Additional, corrected, and/or changed data: b.5., d.7., h.6.
 
Additional Manufacturer Narrative
Device visual evaluation: visual analysis of the photographs a device appears to have the gel appears cloudy in color labeled on the left side.Device analysis performed through photographs, due to the impossibility to perform microscopic analysis it is not possible to determine the most likely failure mode.
 
Event or Problem Description
Healthcare provider reported left side capsular contracture baker grade iii and "gel bleed." device has been explanted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
STYLE 10 SILICONE GEL FILLED BREAST IMPLANT
Common Device Name
PROSTHESIS, BREAST, NONINFLATABLE, INTERNAL, SILICONE GEL-FILLED
Manufacturer (Section D)
ALLERGAN (COSTA RICA)
900 parkway global park
zona franca
CS 
MDR Report Key10148178
Report Number9617229-2020-08704
Device Sequence Number2176848
Product Code FTR
UDI-Device Identifier10888628001299
UDI-Public10888628001299
Combination Product (Y/N)N
Initial Reporter StateLA
PMA/510(K) Number
P020056
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2015
Device Explanted Year2020
Reporter Type Manufacturer
Report Source health professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date (Section B) 08/27/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date12/26/2019
Device Model Number10-450
Device Catalogue Number10-420
Device Lot Number2663934
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 05/14/2020
Supplement Date Received by Manufacturer06/25/2020
07/29/2020
Initial Report FDA Received Date06/12/2020
Supplement Report FDA Received Date07/25/2020
08/27/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age34 YR
-
-