| Model Number |
INTELEPACS INTELEVIEWER 4-10-1 AND UP |
| Medical Device Problem Codes |
Output Problem (3005); No Apparent Adverse Event (3189); Program or Algorithm Execution Failure (4036)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
05/25/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
A hazard may arise when non-dicom compliant jpeg-2000-compressed images are received by the (b)(6) component with dicom bit depth (bits stored tag) that does not match the bit depth in the pixel data codestream.These images are decompressed with a loss of precision in the image data which can lead to failing image registration.
|
| |
|
Search Alerts/Recalls
|