Model Number PLATINIUM DR 1510
Device Problem Failure to Transmit Record (1521)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 05/05/2020
Event Type  malfunction  
Event Description
Reportedly, the subject platinium dr was associated to an home monitor (with serial number (b)(4) that has not sent an alert after the delivery of a shock on (b)(6) 2020, even though "all shocks" alert was activated.On (b)(6) 2020, the device was interrogated by inductive telemetry and it was confirmed the delivery of the shock on (b)(6) 2020.A patient initiated transfer transmission was then performed, to check the proper communication between the subject device and the home monitor.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PLATINIUM
Type of Device
DEFIBRILLATOR, IMPLANTABLE, DUAL-CHAMBER
Manufacturer (Section D)
SORIN GROUP ITALIA S.R.L. - CRM FACILITY
parc d'affaires noveos 4 avenue réaumur
.
clamart 92140
FR  92140
Manufacturer (Section G)
SORIN GROUP ITALIA SRL VIA CRESCENTINO 13040 SALUGGIA ITALY
parc d'affaires noveos 4 avenue réaumur
.
clamart 92140
FR   92140
Manufacturer Contact
laura ouaki
parc d'affaires noveos 4 avenue réaumur
.
clamart 92140
FR   92140
146013429
MDR Report Key10148260
MDR Text Key195435227
Report Number1000165971-2020-00426
Device Sequence Number1
Product Code MRM
UDI-Device Identifier08031527014364
UDI-Public(01)08031527014364(11)180409(17)191109
Combination Product (y/n)N
Reporter Country CodeSP
PMA/PMN Number
P980049
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Reporter Occupation Nurse
Type of Report Initial
Report Date 06/12/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date11/09/2019
Device Model NumberPLATINIUM DR 1510
Device Catalogue NumberPLATINIUM DR 1510
Device Lot NumberS0311
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Distributor Facility Aware Date05/21/2020
Event Location Hospital
Initial Date Manufacturer Received 05/21/2020
Initial Date FDA Received06/12/2020
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured04/09/2018
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-