• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: AVANOS MEDICAL INC. MIC GASTROSTOMY FEEDING TUBE WITH ENFIT CONNECTORS - 18 FR; DH EF BALLOON TUBES PRODUCTS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

AVANOS MEDICAL INC. MIC GASTROSTOMY FEEDING TUBE WITH ENFIT CONNECTORS - 18 FR; DH EF BALLOON TUBES PRODUCTS Back to Search Results
Model Number 8100-18
Medical Device Problem Codes Break (1069); Migration (4003)
Health Effect - Clinical Code No Code Available (3191)
Date of Event 03/17/2020
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
(b)(4).The actual complaint product was not returned for evaluation.A review of the device history record is not possible as no lot number was provided.All information reasonably known as of 11-jun-2020 has been included in this health authority report.Should additional information be obtained, a follow-up health authority report will be provided.The information provided by avanos medical, inc.Represents all of the known information at this time.Avanos medical, inc.Has no independent knowledge of the event reported but is relaying the information that was provided by the user facility where the incident occurred.This product incident is documented in the avanos medical, inc.Complaint database and identified as complaint (b)(4).
 
Event or Problem Description
An fda medwatch / fda user facility report # (b)(4) was received and the following information was provided: "patient had percutaneous endoscopic gastrostomy tube placement.Per the op note: "under endoscopic visualization 3 t-fasteners were placed percutaneously securing the anterior gastric wall against and to abdominal wall.A 1 cm incision was created between these fasteners and carried on down through the rectus fascia.An 18-gauge needle was placed through this incision followed by j-wire using seldinger technique.Under endoscopic visualization a large dilator and sheath was passed over the wire.The wire and the dilator was removed leaving the sheath in place.18 french g-tube was advanced through the sheath and the balloon was inflated with 10 cc of saline and brought back against and to abdominal wall." once he was tolerating tube feedings , he was discharged home three days after.The following morning , patient awoke with abdominal pain.In the ed, ct showed: "g-tube is noted positioned in the stomach.Note is made of punctate extraluminal air to the right of the g-tube tract with small amount of fluid also evident." once admitted to floor , as he was coming out of the bathroom, his g-tube unintentionally fell out of his abd and wound started bleeding.Surgeon consulted and noted: "the patient's peg tube fell out while he was in the bathroom.Reportedly the peg tube was not pulled on.We tested and the balloon was ruptured.The patient had a ct scan of the abdomen earlier today which i personally reviewed which demonstrated a small amount of air near the peg tube which i imagine is just because the balloon was not fully inflated." taken to or that evening, op note states the surgeon located the site of the prior gastrotomy and that "there was staining in the area and it was evident that this had been leaking some around where the tube was.This is a type 4 wound given the obvious hole in the stomach with some contamination." repair to old site completed with graham patch.Then new g-tube placed, stamm type.This surgeon actually put in 2 of these gastrostomy tubes on different patients in the same week.Both of them deflated and/or ruptured, and became dislodged, within 7 days of placement.We have pulled the remainder of this product from our stock for the time being.The surgeon would like to receive a copy of the manufacturer's final report.Thank you.What was the original intended procedure? : g-tube (gastrostomy tube) placed for tube feedings.What problem did the user have (check all that apply) :device failed (e.G.Broke, couldn't get it to wort or stopped working).Date of event: (b)(6) 2020.
 
Additional Manufacturer Narrative
All information reasonably known as of 10-sep-2020 has been included in this health authority report.Should additional information be obtained, a follow-up health authority report will be provided.The information provided by avanos medical, inc.Represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to avanos medical, inc.Avanos medical, inc.Has no independent knowledge of the event reported but is relaying the information that was provided by the user facility where the incident occurred.This product incident is documented in the avanos medical, inc.Complaint database and identified as complaint (b)(4).This information is submitted pursuant to 21cfr803, in compliance with the medical device reporting requirement and should not be considered to be an admission that an avanos medical, inc.Product is defective or caused serious injury.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MIC GASTROSTOMY FEEDING TUBE WITH ENFIT CONNECTORS - 18 FR
Common Device Name
DH EF BALLOON TUBES PRODUCTS
Manufacturer (Section D)
AVANOS MEDICAL INC.
5405 windward parkway
alpharetta GA 30004
MDR Report Key10148482
Report Number9611594-2020-00099
Device Sequence Number2240871
Product Code KNT
UDI-Device Identifier00350770442938
UDI-Public00350770442938
Combination Product (Y/N)N
Initial Reporter StateAZ
Initial Reporter CountryUS
PMA/510(K) Number
K842076
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type User Facility
Report Source health professional,other,use
Initial Reporter Occupation Risk Manager
Type of Report Initial,Followup
Report Date (Section B) 09/10/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number8100-18
Device Catalogue Number104429300
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Initial Date Received by Manufacturer Not provided
Supplement Date Received by Manufacturer08/28/2020
Initial Report FDA Received Date06/12/2020
Supplement Report FDA Received Date09/10/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient Age79 YR
Patient Weight94
-
-