RESPIRONICS CALIFORNIA, INC V60 VENTILATOR; VENTILATOR, CONTINUOUS, MINIMAL VENTILATORY SUPPORT,FACILITY USE
|
Back to Search Results |
|
Model Number V60 |
Device Problem
Incorrect, Inadequate or Imprecise Result or Readings (1535)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Type
malfunction
|
Manufacturer Narrative
|
Date of event: (b)(6) 2020.Date of report: 12jun2020.
|
|
Event Description
|
The field service engineer (fse) started to implement testing, but the unit would not pass the high pressure test.The fse ordered a gas delivery system (gds) to address the reported issue.The fse replaced the gds to address the high pressure test failure.During failure investigation of the returned gds, it was found that the system does not pressurize for leak test.It was also found that the gds failed with o2 device failure and o2 not available.Further investigation of the gds found there to be debris inside the solenoid that was piece from a damaged internal o-ring.The returned gds failed high pressure testing, o2 device failed and o2 not available due to damage to internal o-ring on gds oxygen solenoid valve.Replacing o2 solenoid valve corrected the reported problem.
|
|