| Model Number |
MASK OXYGEN ADULT 3-IN-1 W/ VENT AND TBG |
| Medical Device Problem Code |
Material Separation (1562)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
02/23/2020
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
At this time, the suspect device has not been returned for evaluation.No root cause has been determined at this time, since the investigation is still ongoing.Vyaire medical will submit a supplemental report in accordance with 21 cfr section 803.56 if additional information becomes available.
|
| |
|
Event or Problem Description
|
|
The customer reported that the non-rebreather bag was disconnected from the mask while connected to a patient.The nurse replaced the defective unit with a back up bag.The customer confirmed that there was no patient harm associated with the reported event.
|
| |
|
Additional Manufacturer Narrative
|
|
Result of investigation: the device history record (dhr) review did not show any deviations to manufacturer specification.For functionality failure, it is imperative to receive a sample to duplicate the failure as described.
|
| |
|