• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STRYKER ORTHOPAEDICS-MAHWAH TRIATHLON SYMMETRIC X3 PATELLA; PROSTHESIS, KNEE, PATELLO/FEMOROTIBIAL, SEMI-CONSTRAINED, UNCEMENTED, POROUS, CO

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER ORTHOPAEDICS-MAHWAH TRIATHLON SYMMETRIC X3 PATELLA; PROSTHESIS, KNEE, PATELLO/FEMOROTIBIAL, SEMI-CONSTRAINED, UNCEMENTED, POROUS, CO Back to Search Results
Model Number 5550-G-360
Medical Device Problem Codes Loss of Osseointegration (2408); Osseointegration Problem (3003)
Health Effect - Clinical Codes Pain (1994); Injury (2348); Ambulation or Postural Difficulties (2544)
Date of Event 09/21/2010
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Review of the device history records indicate devices were manufactured and accepted into final stock with no relevant reported discrepancies.There have been no other similar events for the lot referenced.It was noted that the device is not available for evaluation.Should additional information become available, it will be provided in a supplemental report upon completion of the investigation.
 
Event or Problem Description
Information received in follow-up includes an operative report for a revision surgery performed (b)(6) 2010.A synovectomy and removal of a loose patellar component was performed.A new patella was not implanted.
 
Event or Problem Description
Information received in follow-up includes an operative report for a revision surgery performed (b)(6) 2010.A synovectomy and removal of a loose patellar component was performed.A new patella was not implanted.
 
Additional Manufacturer Narrative
Reported event: an event regarding revision due to loosening and failure of patellar component is reported involving triathlon patella.The event was confirmed for failure of patellar component based on medical review.Method & results: product evaluation and results- product inspection - not performed because device was not returned.Clinical review: the following adverse events were identified: left: failed patellar component with resection arthroplasty, revision left knee, periprosthetic infection, multiple aspirations and infection.Right: patellar component failure and revision, 2nd revision for knee dislocation and patellar tendon rupture.Conclusion of assessment: the mechanisms of failure and causes of revision in both knees appeared dynamic and unrelated to the implants.This was a very complicated case in a patient with multiple comorbid conditions.-product history review: indicated all devices were manufactured and accepted into final stock with no reported discrepancies.-complaint history review: there have been no other similar events for the lot referenced.Conclusions: the event was reported about revision due to loosening and failure of patellar component involving triathlon patella.The event was confirmed for failure of patellar component based on medical review.The exact cause of the event could not be determined because insufficient information was provided.Further information is needed to complete the investigation for determining its root cause.No further investigation for this event is possible at this time.If additional information become available to indicate further evaluation is warranted, this record will be reopened.H3 other text : device not returned.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TRIATHLON SYMMETRIC X3 PATELLA
Common Device Name
PROSTHESIS, KNEE, PATELLO/FEMOROTIBIAL, SEMI-CONSTRAINED, UNCEMENTED, POROUS, CO
Manufacturer (Section D)
STRYKER ORTHOPAEDICS-MAHWAH
325 corporate drive
mahwah NJ 07430
MDR Report Key10152056
Report Number0002249697-2020-01169
Device Sequence Number13960126
Product Code MBH
UDI-Device Identifier07613327014778
UDI-Public07613327014778
Combination Product (Y/N)N
Initial Reporter StateKY
Initial Reporter CountryUS
PMA/510(K) Number
K141056
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2008
Device Explanted Year2010
Reporter Type Manufacturer
Report Source health professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 08/27/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date07/31/2013
Device Model Number5550-G-360
Device Catalogue Number5550-G-360
Device Lot Number779M
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 05/26/2020
Supplement Date Received by Manufacturer08/04/2020
Initial Report FDA Received Date06/15/2020
Supplement Report FDA Received Date08/27/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
Patient Age51 YR
-
-